① Handpicece port:connect the handpiece to console.
② Pedal switch port:connect the pedal switch to console.
③ Power socket:connect the console to power source with power cable.
④ Leakage protection switch:turn on or turn off the console.
5 Method for loading and unloading the punch
1) Rotate the fixing tube of handpiece in one direction,see Fig-1.
2) Insert the punch into the central hole of handpiece,
3) Rotate the fixing tube in reverse direction,the punch is loaded. see Fig-2.
4) Unload the punch in reverse process.

Operating method
preparation
1) check the punch,if not intact,take a new one.
2) read and comprehend the user manual.
Operating procedure
1) Connect the console to power source,
2) Turn on the switch,
3) Choose bidirectional or monodirectional mode,
4) Set the parameter of speed and angle,
5) Press the count button and switch to the count interface,
6) lay up the punch near the scalp for hair grafts extraction,
7) puncture in the scalp,
8) pick up the hair grafts with forcep,
9) Implant the hair grafts in bald area.
Sterilization
|
Condition
|
Suitable components
|
Autoclaving
|
temperature:120℃
duration:30 minutes
|
punch
|
EO fumigation
|
EO concentration:600 (±30) ppm
temperature:50(+/-3)℃
RH:50(+/-5)%
duration:6 hours
gas composition:EO 30% + CO2 70%
pressure:0.5 - 0.7kg/cm2
|
Punch、handpiece
|